Catalog Number 1011487-15
Medical Device Problem Codes Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that the procedure on (b)(6) 2023 was to treat a 70% stenosed lesion in the left vertebral artery ostium.A guide wire was placed and pre-dilatation was performed.The 4.0x15mm herculink elite stent delivery system (sds) was attempted to be advanced to the target lesion; however, failed to advance due to the anatomy.When the sds was being removed to perform additional pre-dilatation, the stent dislodged at the ostium of the vertebral artery.The herculink sds was removed and a 1.5x15mm balloon was used to attempt to remove the stent.The stent was retracted back to the femoral artery, but was not able to be removed with the balloon.The patient had no discomfort and blood flow was smooth.The patient was placed temporarily on anti-platelet medication and monitored.Then, on 10/03/2023 digital subtraction angiography (dsa) was performed and a surgical incision was made to remove the stent.The patient is in good condition.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported difficulties and subsequent surgical intervention, unexpected medical intervention (additional therapy/non-surgical treatment), and medication required appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Common Device Name
STENT, RENAL
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18132327
Report Number2024168-2023-12619
Device Sequence Number20771593
Product Code NIN
UDI-Device Identifier08717648077838
UDI-Public08717648077838
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P110001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Catalogue Number1011487-15
Device Lot Number2041161
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/23/2023
Supplement Date Received by Manufacturer01/10/2024
Initial Report FDA Received Date11/14/2023
Supplement Report FDA Received Date01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-