Returned qty.(b)(4) inserts, 1.80395, 23139-00096930, kd, beyond useful life, test method / gp005-wi02, inductance: gauge id#: 5349, due: 11/30/2024, result: 3.03, meets spec / does not meet spec: n/a, tip condition: meets spec / does not meet spec: n/a, stack condition: meets spec / does not meet spec: n/a, tested on: g136 gage id#: 9626-2, due date: 03/31/2024, set pressure: 35 psi, water flow: output rate: 35 psi, meets spec / does not meet spec: n/a, spray pattern: meets spec / does not meet spec: n/a, water leak: hp sealing ring, proximal ring, distal ring, seam leak: n/a, vibration: meets spec / does not meet spec: n/a, grip condition: meets spec / does not meet spec: n/a, other visual observation: insert tip is bent.Also insert was tested on a digital thermometer i.D.#: 6116a, due date: 09/30/2024.The temperature was 87.4° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed / not confirmed.2.80395, 23139-00096930, kd, beyond useful life, test method / gp005-wi02, inductance: gauge id#: 5349, due: 11/30/2024, result: 3.06, meets spec / does not meet spec: n/a, tip condition: meets spec / does not meet spec: n/a, stack condition: meets spec / does not meet spec: n/a, tested on: g136 gage id#: 9626-2, due date: 03/31/2024, set pressure: 35 psi, water flow: output rate: 35 psi, meets spec / does not meet spec: n/a, spray pattern: meets spec / does not meet spec: n/a, water leak: hp sealing ring, proximal ring, distal ring, seam leak: n/a, vibration: meets spec / does not meet spec: n/a, grip condition: meets spec / does not meet spec: n/a.Other visual observation: insert tip is good, no fault found.Also insert was tested on a digital thermometer i.D.#: 6116a, due date: 09/30/2024.The temperature was 85.5° f, no fault found.The specification states are not to exceed 118.4°f.(per frs-9175).Alleged event: confirmed / not confirmed.
|