• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. PHILIPS CPAP DEVICE; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. PHILIPS CPAP DEVICE; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Cough (4457); Unspecified Respiratory Problem (4464)
Date of Event 08/13/2021
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap and mechanical ventilator devices.The manufacturer received information alleging an issue related to a philips cpap device's sound abatement foam.The patient has alleged hypersensitive mucus membranes with a lot of coughing, mucus, dry cough and also the device's foam degradation.No medical intervention was required by the patient.There was no report of serious or permanent harm or injury.The device has not yet been returned to the manufacturer for investigation.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
Since no device information was provided, the exact recall number is unknown.Possible recall numbers include z-1972, z-1973, and z-1974.H3 other text : not returned to manufacturer.
 
Additional Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap and mechanical ventilator devices.The manufacturer received information alleging an issue related to a philips cpap device's sound abatement foam.The patient has alleged hypersensitive mucus membranes with a lot of coughing, mucus, dry cough and also the device's foam degradation.No medical intervention was required by the patient.There was no report of serious or permanent harm or injury.The device has not yet been returned to the manufacturer for investigation.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.Udi related data quality updates only.The udi in section d4 was previously reported in the incorrect format.The udi information has been updated to the correct format.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHILIPS CPAP DEVICE
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18133168
Report Number2518422-2023-29860
Device Sequence Number12710282
Product Code BZD
UDI-Device Identifier00606959027220
UDI-Public(01)00606959027220
Combination Product (Y/N)N
Initial Reporter CountryDA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/13/2021
Supplement Date Received by Manufacturer08/13/2021
Initial Report FDA Received Date11/14/2023
Supplement Report FDA Received Date08/13/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/16/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-