| Model Number |
0998-00-0800-32 |
| Medical Device Problem Codes |
Break (1069); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when checking the cardiosave intra-aortic balloon pump (iabp) with a non-sterile demo catheter, the adapter for the catheter extension was previously loosened with a clamp when trying to unscrew it.The pneumatic module was destroyed.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
A getinge field service engineer (fse) evaluated the unit for the reported malfunction.The fse stated that the pneumatic module was destroyed in the clinic due to incorrect handling.The device had not yet been in use.The fse replaced the pneumatic module assembly (0997-00-1178).The fse performed maintenance and the all manifold test.Various checks and calibrations were performed.The unit passed all performance and safety tests according to factory specifications.The iabp was returned to the customer and released for clinical use.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only** providing updated device identification information in alignment with gudid.
|
| |
|