• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. CARDIOSAVE HYBRID TYPE J PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. CARDIOSAVE HYBRID TYPE J PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-32
Medical Device Problem Codes Break (1069); Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when checking the cardiosave intra-aortic balloon pump (iabp) with a non-sterile demo catheter, the adapter for the catheter extension was previously loosened with a clamp when trying to unscrew it.The pneumatic module was destroyed.There was no patient involvement.
 
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Additional Manufacturer Narrative
A getinge field service engineer (fse) evaluated the unit for the reported malfunction.The fse stated that the pneumatic module was destroyed in the clinic due to incorrect handling.The device had not yet been in use.The fse replaced the pneumatic module assembly (0997-00-1178).The fse performed maintenance and the all manifold test.Various checks and calibrations were performed.The unit passed all performance and safety tests according to factory specifications.The iabp was returned to the customer and released for clinical use.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
**udi related data quality updates only** providing updated device identification information in alignment with gudid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE J PLUG
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP.
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key18133266
Report Number2249723-2023-04859
Device Sequence Number18037708
Product Code DSP
UDI-Device Identifier10607567111117
UDI-Public(01)10607567111117(21)CB356764G2
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K112372
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-0800-32
Device Catalogue Number0998-00-0800-32
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/03/2023
Supplement Date Received by Manufacturer07/22/2024
10/01/2024
Initial Report FDA Received Date11/14/2023
Supplement Report FDA Received Date07/22/2024
10/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2022
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-