WILSON-COOK MEDICAL INC ACUSNARE POLYPECTOMY SNARE SOFT; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
| Catalog Number |
SASM-1-S |
| Medical Device Problem Code |
Off-Label Use (1494)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the additional information provided by the user stated that a cold polypectomy was performed.This is against the intended use of the device.The intended use for this device states: "this device is used with an electrosurgical unit for endoscopy polypectomy.Do not use this device for any purpose other than the stated intended use." this is the most likely cause of the report.Prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered remote.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.Additional comments regarding this report: based on the information provided that a cold polypectomy was performed, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|
| |
|
Event or Problem Description
|
|
During a colonoscopy, the physician used a cook acusnare polypectomy snare soft.It was reported [that] the nurse stated that, when she was closing the snare to snip off the polyp, when she was about half way closed, she felt like there was a drag or delay resistance in the tension on the wire.The procedure was successfully completed with the device in question.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|
|
|