AVANOS MEDICAL INC. CORTRAK 2 NG/NI FEED TUBE, ELECTROMAG STYLET, ENFIT; DH CORTRAK DISPOSABLES
|
Back to Search Results |
|
| Model Number |
40-9368TRAK2 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
10/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual complaint product was not returned for evaluation.A review of the device history record is in-progress.All information reasonably known as of 14-nov-2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|
Event or Problem Description
|
|
It was reported via fda user facility report mw report mw5146866 the following information: on (b)(6) 2023, the patient's nasal duodenal (nd) tube clogged and despite multiple attempts to flush and use clot-buster medications, obstruction was not cleared.When nd was pulled, the distal portion broke off and remains in patient's esophagus.Gi was consulted and patient was taken to operating room on (b)(6) 2023 to have surgically removed.Portion of the product was saved.Additional information received 01-nov-2023 stating the patient returned to (operating room) or for removal of the device.The device was initially inserted on (b)(6) 2023.The patient is currently deceased, no additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record for the reported lot number, 30243722, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.All information reasonably known as of 17-jun-2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
|
| |
|