| Model Number |
26280 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Failure to Advance (2524); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1.Initial reporter phone: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the stent was damaged and difficult to reconstrained.The 80% stenosed target lesion was located in the severely tortuous and non-calcified iliac vein.Pre-dilatation was performed with a non-boston scientific (bsc) balloon, and a 12x90/8fr uni plus halo 75cm wallstent endoprosthesis was advanced but failed to cross the lesion along with a non-bsc sheath.However, angiography showed the tip of the stent was exposed and partially damaged.In addition, the stent was inadequacy reconstrained during withdrawal.The device was simply removed as a whole together and the procedure was completed with another of the same device with no patient complications reported and the patient status was stable.
|
| |
|