Model Number LF1823
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during a procedure, the handpiece jaw no longer open or close after an hour.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted intermittently, the lever did not latch when closing the jaws.The device was activated multiple times on a saline soaked gauze pad with satisfactory results.All seal cycles were completed satisfactorily and end tones were heard indicating completed activation cycles.No electrical or wiring issues were found.It was reported that the device was difficult and impossible to close and jaws were difficult to open.The reported issue was confirmed.The most likely cause was determined to be manufacturing related.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18136169
Report Number1717344-2023-01403
Device Sequence Number12762194
Product Code GEI
UDI-Device Identifier10884521705876
UDI-Public10884521705876
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K162941
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLF1823
Device Catalogue NumberLF1823
Device Lot Number31370359X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2023
Supplement Date Received by Manufacturer02/19/2024
Initial Report FDA Received Date11/14/2023
Supplement Report FDA Received Date02/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-