• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SILK ROAD MEDICAL ENROUTE TRANSCAROTID STENT SYSTEM; ENROUTE TSS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SILK ROAD MEDICAL ENROUTE TRANSCAROTID STENT SYSTEM; ENROUTE TSS Back to Search Results
Model Number SR-0840-CS
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Obstruction/Occlusion (2422); Ischemia Stroke (4418); Thrombosis/Thrombus (4440)
Date of Event 10/19/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product associated with this complaint was not returned to the manufacturer for analysis.A review of the manufacturing records for this device was completed and no issues were identified that could have led to the adverse event reported.There is no indication that a malfunction of the srm device occurred.At this time, it is unknown if the reported event is related to medication/patient response (plavix resistance), a procedural issue, or a silk road medical device failure, hence, the event will be reported out of abundance of caution.Complaints will continue to be reviewed and monitored for trends.All reasonably available information has been provided by the company at the time of submission of this report.The fields that are blank are not an omission and indicate that the information is either not applicable or currently unavailable.If additional information is received, a supplemental report will be filed.
 
Event or Problem Description
It was reported that hours after the successful completion of a transcarotid artery revascularization (tcar) procedure, the patient experienced stroke symptoms and the stent had occluded.The physician re-ballooned the stent and performed a mechanical thrombectomy.Additional information indicated that the patient's symptoms were resolving.At this time, it is unknown if the reported event is related to medication/patient response (plavix resistance), a procedural issue, or a silk road medical device failure, hence, the event will be reported out of abundance of caution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENROUTE TRANSCAROTID STENT SYSTEM
Common Device Name
ENROUTE TSS
Manufacturer (Section D)
SILK ROAD MEDICAL
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer (Section G)
SILK ROAD MEDICAL
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer Contact
shelrin devi
1213 innsbruck drive
sunnyvale, CA 94089
4087209002
MDR Report Key18136736
Report Number3014526664-2023-00190
Device Sequence Number6588647
Product Code NIM
UDI-Device Identifier00811311020478
UDI-Public(01)00811311020478(17)260228(10)18191361
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P140026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberSR-0840-CS
Device Catalogue NumberSR-0840-CS
Device Lot Number18191361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/21/2023
Initial Report FDA Received Date11/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient SexMale
-
-