The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb blackout.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.In addition, our technician confirmed that the objective lens dusty, the bending rubber leak, the remote control buttons leak, the bending rubber cut, the remote control buttons cut, the electrical pin connector dirty, the ccd module with drive pcb defective, the ccd drive pcb defective, and the electrical pin connector defective; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|