|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling and applicable manufacture records.Based on a review of the information, the following was concluded: the device was returned to service facility for evaluation.During evaluation, the reported issue of the unit does not charge and shuts down was confirmed.The prevue fails to charge and shuts off abruptly when disconnected from ac power.The root cause is due to an internal failure within the lithium-ion battery.H3 other text : evaluation findings are in section h11.
|