| Model Number |
802018 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The fsr replaced the flow module.Mechanical, electrical and visual testing was performed.The unit operated to the manufacturer's specifications.
|
| |
|
Event or Problem Description
|
|
During field installation of the device, the field service representative (fsr) reported that the flow module failed testing.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
During laboratory evaluation, the product surveillance technician (pst) connected the flow pod to lab use only (luo) testing equipment.There was no flow displayed on the screen when the pod was in use.The information on screen was as if there was no flow sensor attached.There were dashes displayed for the flow reading and a 'check sensor' message was displayed.With a luo generic printed circuit board (pcb) installed into the flow pod, the issue remained.With a luo application board installed into the flow pod, the flow displayed on the screen and the issue was corrected.The application pcb was the root cause of the reported complaint.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|