• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CURONIX LLC FREEDOM SPINAL CORD STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CURONIX LLC FREEDOM SPINAL CORD STIMULATOR Back to Search Results
Model Number FR8A-RCV-A0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Discomfort (2330); Numbness (2415); Swelling/ Edema (4577)
Date of Event 10/18/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The unintended stimulation/new pain questionnaire was reviewed for potential causes of the reported issue.Based on this review, the waa providing therapy when new pain occurred, changing the waa parameters intentionally or unintentionally, the patient having a non-curonix device implanted, not performing x-rays before and after the stimulation was experienced, device migration, and the patient experiencing new pain in a new area that is not targeted for therapy have been ruled out as potential causes.However, the patient has spinal stenosis and the implanting clinician believes the numbness was due to the lead placement.The stimulator is used to treat pain.The cause of the reported issue is due to patient being a poor candidate due to health, weight, age, or mental capacity as the patient has spinal stenosis (user error-clinician).Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in unintended stimulation/new pain issues.Unintended stimulation/new pain issue rates remain acceptably low; thus, capa is not required. unintended stimulation/new pain issue rates will continue to be tracked and trended.
 
Event or Problem Description
The patient reported numbness, immobility, discomfort, and severe swelling in their leg.An explant procedure was performed on (b)(6) 2023 which resolved the swelling.The patient is doing well, mobility has returned, and they are expected to be implanted with a new lead, however, a date has not been set.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEDOM SPINAL CORD STIMULATOR
Common Device Name
SPINAL CORD STIMULATOR
Manufacturer (Section D)
CURONIX LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer (Section G)
CURONIX LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer Contact
alicia pagliaro
1310 park central boulevard s.
pompano beach, FL 33064
8009655134
MDR Report Key18143771
Report Number3010676138-2023-00234
Device Sequence Number17889441
Product Code GZB
UDI-Device Identifier00818225020693
UDI-Public(01)00818225020693(17)240401
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K182720
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date04/01/2024
Device Model NumberFR8A-RCV-A0
Device Lot NumberSWO220428
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/18/2023
Initial Report FDA Received Date11/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexMale
-
-