• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NELLCOR PURITAN BENNETT MEXICO SA NELLCOR; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NELLCOR PURITAN BENNETT MEXICO SA NELLCOR; OXIMETER Back to Search Results
Model Number MAXFAST
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Blister (4537)
Date of Event 04/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during use, the sensor produced a pressure injury sustained on the occiput (back of the head) from the strap of the sensor used at the back of the head.It was advised that the pressure injuries were caused by a combination of the sensor headband, patient condition and pressure injury prevention measures.Measures have since been taken to reduce the risk and the removal of the sensor was one of those measures.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Additional information: b5, g3 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during use, the sensor produced a pressure injury sustained on the occiput (back of the head) from the strap of the sensor used at the back of the head.It was advised that the pressure injuries were caused by a combination of the sensor headband, patient condition and pressure injury prevention measures.Measures have since been taken to reduce the risk and the removal of the sensor was one of those measures.The sensor was placed on the forehead and nil additional forces.Site was checked periodically with rolls and removed when the pressure injury identified.The pressure injuries did not contributed to the death.The patient died.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The involved device was not returned.A comprehensive exam ination could not be performed, because the returned sample was not received in a state that allowed full functional or visual assessment.It was reported that the sensor produced a pressure injury.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the sensor produced a pressure injury sustained on the occiput (back of the head) from the strap of the sensor used at the back of the head.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
OXIMETER
Manufacturer (Section D)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX  22225
Manufacturer (Section G)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX   22225
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18144848
Report Number1282497-2023-00059
Device Sequence Number15329209
Product Code DQA
UDI-Device Identifier10884522039710
UDI-Public10884522039710
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K012891
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberMAXFAST
Device Catalogue NumberMAXFAST
Device Lot Number220660192H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2023
Supplement Date Received by Manufacturer02/28/2024
07/05/2024
10/28/2024
Initial Report FDA Received Date11/15/2023
Supplement Report FDA Received Date03/22/2024
07/10/2024
10/29/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-