• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS CORPORATION TURBO-ELITE LASER ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS CORPORATION TURBO-ELITE LASER ATHERECTOMY CATHETER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 410-154
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Unintended Radiation Exposure (4565)
Date of Event 09/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
A peripheral atherectomy procedure commenced, accessing the lesion from the femoral and brachial arteries, using a contralateral approach (lesion on opposite side of the body from the access site).A spectranetics turbo-elite laser atherectomy catheter was used to treat the patient.When removing the turbo-elite from the body, it appeared stuck on the introducer sheath, which was positioned across the curve in the vasculature, near the femoral site.However, after a few seconds of pulling, the device was successfully removed and appeared intact.The procedure was completed with no reported patient harm.Based upon the device evaluation, a breach in the outer jacket, along with exposed fibers in the area, was discovered.This event is being reported for unintended radiation exposure, potential for harm.
 
Additional Manufacturer Narrative
A2): patient''s date of birth, age unk.A3): patient''s gender unk.A4): patient''s weight unk.A5a./5b.): patient''s ethnicity/race unk.B6): relevant tests/laboratory data unk.B7): other relevant history unk.H3): visual inspection of the turbo-elite found a breach to the outer jacket, 6.5 cm from the distal tip.The outer jacket and inner lumen were wrinkled and fractured at the area of the breach, pulled away approximately 2 cm, exposing the fibers in the area.No broken fibers were detected.H6): based on the device evaluation and investigation, the cause of the damage cannot be established.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBO-ELITE LASER ATHERECTOMY CATHETER
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
719447-2465
MDR Report Key18144903
Report Number3007284006-2023-00064
Device Sequence Number15338698
Product Code MCW
UDI-Device Identifier00813132024703
UDI-Public(01)00813132024703(17)250613(10)FBE23E25A
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K170059
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/13/2025
Device Model Number410-154
Device Catalogue Number410-154
Device Lot NumberFBE23E25A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDE WIRE MANUFACTURER/SIZE UNK; INTRODUCER SHEATH MANUFACTURER/SIZE UNK; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM
Patient SexUnknown
-
-