| Catalog Number |
ASHA4830-01 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament reconstruction surgery, the electrode of the ambient super multivac wand detached from the tip and fell off inside the patient.It was removed with suction and tweezers.The procedure was completed with non-significant delay using a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.Visual inspection observed the returned instrument shows no manufacturing abnormalities.The electrode has been used and is still attached to the tip.The electrode is intact with no missing pieces.Bio debris is present.Product was out of the original packaging.No packaging returned.A functional evaluation found the device displayed settings (1,7) when plugged in.Plasma and coagulation were generated as intended.The resistance measured at 1.61 kilo ohms.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause could not be determined since the reported malfunction could not be duplicated during the investigation.Factors that can contribute to the reported event include hitting the tip against a hard surface such as a metal or bone, mechanical displacement of the tip through applied force or an impact event, or using the wand above default output settings causing excessive electrode erosion.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|