| Model Number |
MH-946 |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No device was returned to olympus for evaluation.However, as part of the investigation into this report, an olympus endoscopy support specialist (ess) has been dispatched on-site to assess the reprocessing practices at the user facility.
|
| |
|
Event or Problem Description
|
|
Olympus was informed that during an onsite visit, the user facility staff was not reprocessing the injection tube after each use.No patient infections or injuries reported.
|
| |
|
Event or Problem Description
|
|
The customer requested an olympus endoscopy support specialist conduct a reprocessing in-service.While onsite, it was discovered the customer was not reprocessing the auxiliary water tube, suction cleaning adapter, and injection tube after each use.There was no report of patient harm associated with this event.Related patient identifiers: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Correction to b5 and h10 for information inadvertently left out of previous report.Three attempts were performed to obtain additional information, but no response was received from the customer.Olympus endoscopy support specialist while on-site reviewed reprocessing of a colonovideoscope, auxiliary water tube, suction cleaning adapter, and injection tube according to the reprocessing manual.It was recommended to reprocess the auxiliary water tube, suction cleaning adapter, and injection tube according to the manual.The options of soaking in high level disinfectant or steam sterilization after each use were discussed.The device history record was unable to be reviewed for this device since the lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, it is likely there was a difference in recognition on item handling or reprocessing steps between olympus recommendation and the user facility.Olympus expert staff has already conducted training on proper item handling for the user facility.Preventive measure is described in evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
|
| |
|