| Catalog Number |
2112550 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: brief inquiry description: it was reported that after initial setup the apex transfer set had a black particulate in port 2 macro of the tubing just before the manifold.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample without packaging was returned to the manufacturer for evaluation.Through visual examination, an embedded particulate on the outside of the tube that inserts into the manifold was observed.The sample is not within specification; the reported defect is confirmed.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Incidents of this nature are attributed to operator oversight during the assembly of the product.As a result of this occurrence, a formal awareness training session was conducted with all applicable personnel involved in the packaging and inspection of this product.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|