• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977762
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare professional reported with a description of loading area of the cartridge was small, making it difficult for forceps to push in, and the tip was small, making it difficult for intraocular lens (iol) to be injected.There was no patient harm.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
One unused company cartridges was returned in the opened pouch.The unused company cartridge was microscopically examined with no damage or abnormalities observed.The loading area and the cartridge tip were measured and verified to meet specifications.The associated products were not provided.It is unknown if qualified associated products were used.No problem was found with the returned unused company cartridge.The cartridge was measured and verified to meet specifications.Cartridge manufacturing is a validated operation, with specifications that are maintained and documented.No changes have been made for the manufacturing of cartridges.Supplier manufacturing records as well as their certificate of compliance verify the dimensional measurements are within specifications.It is unknown if qualified associated products were used.The ifu instructs: company foldable iols are qualified for use with company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C
Common Device Name
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18148555
Report Number1119421-2023-01911
Device Sequence Number12509630
Product Code MSS
UDI-Device Identifier00380659777622
UDI-Public00380659777622
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K112977
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/28/2026
Device Catalogue Number8065977762
Device Lot Number15617227
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/23/2023
Supplement Date Received by Manufacturer12/18/2023
Initial Report FDA Received Date11/16/2023
Supplement Report FDA Received Date01/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/30/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-