| Model Number |
806455 |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during use of the device for cardiopulmonary bypass (cpb), the occluder lost calibration and needed to be recalibrated.There were no reported adverse consequences to the patient.No other details regarding the nature of this event were provided.
|
| |
|
Event or Problem Description
|
|
Per clinical review: the user reported an issue with the heart lung machine (hlm) during cardiopulmonary bypass (cpb).Specifically, the user stated they lost calibration of the occluder and the system was requiring a recalibration.They decided to not use the occluder the rest of the case.The user reported no delay to the case, the surgery was completed successfully with no patient harm nor blood loss.
|
| |
|
Additional Manufacturer Narrative
|
|
The field service representative could not duplicate the reported issue.He performed release testing on the occluder and no issues were found.The unit operated to the manufacturer's specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint could not be confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|