MAUDE Adverse Event Report: CONSOLIDATED MEDICAL EQUIPMENT COMPANY AIRSEAL 12/120MM LPI PORT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONSOLIDATED MEDICAL EQUIPMENT COMPANY AIRSEAL 12/120MM LPI PORT; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number IAS12-120LPI
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Air Embolism (1697)
Date of Event 10/23/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Conmed japan reported on behalf of the customer that ias12-120lpi, airseal 12/120mm lpi port device was being used during a robot-assisted nephrectomy procedure post operatively on (b)(6) 2023 when it was reported ¿subcutaneous emphysema occurred, although the causal relationship with the product is unknown.The patient was not required to undergo any additional treatment, and the patient was merely observed.The confirmed event was that gas leakage occurred during the surgery, and pneumoperitoneum could not be maintained.After replacing the trocar and tube, pneumoperitoneum was restarted.After the surgery, doctors noticed that she had developed subcutaneous emphysema.The presence or absence of product defects and the causal relationship are unknown.The facility believes that if there is a hole somewhere in the trocar, it could lead to subcutaneous emphysema from the abdominal wall, so the actual item needs to be inspected and reported.¿.There was no report of medical intervention, or extended hospitalization to the patient or user.The device asm-evac1, trilumen filtered tube set was also used in this procedure and will be listed as a concomitant device.This report is being raised on the reported injury due to the patient being diagnosed with subcutaneous emphysema.
 
Additional Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Additional Manufacturer Narrative
Received one ias12-120lpi in opened original packaging.Lot number was verified.Performed a visual inspection, there were no obvious signs of abnormalities or defects.Performed a functional inspection using the airseal i.F.S.(c8457), the device functioned as intended and was able to maintain pneumoperitoneum.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.(b)(4).Per the instructions for use, the user is advised that higher insufflation pressures (> 15 mm hg) of carbon dioxide insufflation can increase the risk of hypercarbia, subcutaneous emphysema, pneumomediastinum, pneumothorax, pneumo-scrotum, and urinary retention.Use extreme caution during airseal access port insertion.Improper use of this product can result in life-threatening injury to internal organs and vessels.Ensure that adequate pneumoperitoneum or pneumo-rectum is established.Ensure that the patient is properly positioned so that organs are away from the penetration site.Direct the airseal access port¿s tip away from significant vessels and organs.Do not use excessive downward force.We will continue to monitor for trends through the complaint system to assure patient safety.
 
Event or Problem Description
Conmed japan reported on behalf of the customer that ias12-120lpi, airseal 12/120mm lpi port device was being used during a robot-assisted nephrectomy procedure post operatively on (b)(6) 2023 when it was reported ¿subcutaneous emphysema occurred, although the causal relationship with the product is unknown.The patient was not required to undergo any additional treatment, and the patient was merely observed.The confirmed event was that gas leakage occurred during the surgery, and pneumoperitoneum could not be maintained.After replacing the trocar and tube, pneumoperitoneum was restarted.After the surgery, doctors noticed that she had developed subcutaneous emphysema.The presence or absence of product defects and the causal relationship are unknown.The facility believes that if there is a hole somewhere in the trocar, it could lead to subcutaneous emphysema from the abdominal wall, so the actual item needs to be inspected and reported.¿.There was no report of medical intervention, or extended hospitalization to the patient or user.The device asm-evac1, trilumen filtered tube set was also used in this procedure and will be listed as a concomitant device.This report is being raised on the reported injury due to the patient being diagnosed with subcutaneous emphysema.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRSEAL 12/120MM LPI PORT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX  31136
Manufacturer (Section G)
CONSOLIDATED MEDICAL EQUIPMENT COMPANY
ave. alejandro dumas 11321
complejo industrial chihuahua
chihuahua 31136
MX   31136
Manufacturer Contact
samantha dewberry
525 french road
utica, NY 13502
3152230184
MDR Report Key18150514
Report Number3007305485-2023-00240
Device Sequence Number6721064
Product Code GCJ
UDI-Device Identifier10845854045688
UDI-Public(01)10845854045688(17)250719(10)202207204
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K190303
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/19/2025
Device Catalogue NumberIAS12-120LPI
Device Lot Number202207204
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2023
Supplement Date Received by Manufacturer12/19/2023
Initial Report FDA Received Date11/16/2023
Supplement Report FDA Received Date12/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASM-EVAC1, TRILUMEN FILTERED TUBE SET; ASM-EVAC1, TRILUMEN FILTERED TUBE SET
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-