Catalog Number 91-9513SC / 519270
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code Sensitivity of Teeth (2427)
Date of Event 01/30/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
 
Event or Problem Description
According to the available information the device was explanted and replaced due to not achieving sufficient erection.
 
Event or Problem Description
According to the available information the device was explanted and replaced due to not achieving sufficient erection.
 
Additional Manufacturer Narrative
No functional abnormalities were noted with either rod 1 or rod 2.The information received indicated the device had a malfunction, but because no functional abnormalities were noted with the returned components, the complaint could not be confirmed as reported.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENESIS MALLEABLE PENILE PROSTHESIS
Common Device Name
PENILE PROSTHESIS, ROD/CYLINDER
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
MINNEAPOLIS
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA   3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key18151184
Report Number2125050-2023-01417
Device Sequence Number5788035
Product Code FAE
UDI-Device Identifier05708932324809
UDI-Public05708932324809
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K040959
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number91-9513SC / 519270
Device Lot Number8658506_5192702022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/16/2023
Supplement Report FDA Received Date01/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient SexMale
-
-