|
This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the investigation, the nozzle was adhered/clogged with foreign material.Which was believed, to be chemical residue (detergent).However, further detail of the foreign material is unknown.The likely cause of the event, is due to leakage occurred, at biopsy channel.Therefore, the root cause of the reported event is unable to be determined.The event can be detected and prevented, by following the instructions for use (ifu) sections: ifu states, the detection method in gif/cf/pcf-190 series, operation manual chapter 3: preparation and inspection.Ifu states, the preventative measures in gif/cf/pcf-190 series, reprocessing manual chapter 5: reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
|