• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. POD PACKING COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. POD PACKING COIL; HCG, KRD Back to Search Results
Catalog Number RBYPODJ5
Medical Device Problem Codes Physical Resistance/Sticking (4012); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 1.3005168196-2023-00511, 2.3005168196-2023-00513, 3.3005168196-2023-00514.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a pulmonary arteriovenous malformation (avm) using a pod packing coils (pod pc), a non-penumbra catheter, a non-penumbra microcatheter, and a lantern delivery microcatheter (lantern).It was noted that the patient''s anatomy was tortuous.During the procedure, the physician experienced resistance while advancing a pod pc (b)(4) through the non-penumbra microcatheter.It was reported that the pod pc was unable to reach the target vessel and unintentionally detached within the microcatheter.Therefore, the microcatheter containing the detached pod pc was removed, and the pod pc was removed from the microcatheter.The physician then reinserted the microcatheter into the patient and attempted to advance another pod pc (b)(4); however, the same issue occurred.Therefore, the detached pod pc and microcatheter were removed together.The physician then inserted a lantern into the patient and attempted to advance a pod pc (b)(4) through the lantern.However, resistance was encountered, and the pod pc unintentionally detached inside the lantern.Therefore, the lantern containing the detached pod pc was removed, and the pod pc was removed from the microcatheter.The physician then reinserted the lantern into the patient and attempted to advance another pod pc (b)(4); however, the same issue occurred.Therefore, the detached pod pc and lantern were removed together.The procedure was completed using a ruby coil and another microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD PACKING COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key18151611
Report Number3005168196-2023-00512
Device Sequence Number5709455
Product Code HCG
UDI-Device Identifier00814548017631
UDI-Public814548017631
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K170852
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/16/2023
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBYPODJ5
Device Lot NumberF00000758
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2023
Initial Report FDA Received Date11/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-