• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH ARTIS ZEE FLOOR; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH ARTIS ZEE FLOOR; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10094135
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2023
Type of Reportable Event Malfunction
Event or Problem Description
Siemens became aware of an incident that occurred while operating the artis zee floor system.During a patient procedure, the physician repositioned the radiation protection screen, and it was hit by the c-arm.This resulted in a part of the lead screen breaking off.Additionally provided information stated that the broken part of the lead screen was caught by the plastic covering of the radiation protection screen.We have no indications of any adverse effects on the health status of the involved patient.
 
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported event.A supplemental report will be filed if additional information becomes available.
 
Additional Manufacturer Narrative
The investigation of the reported incident did not reveal any indication of an unintended system movement.The artis zee floor system did not cause or contribute to the collision.There is no indication for a system malfunction.The described collision situation may only occur if the user does not check for surroundings and potential obstacles prior to releasing system motions.The operator manual contains adequate instructions about system movement and how the operator can avoid possible collision.The damaged live monitor was exchanged as part of the service activity.The concerned system works as specified.The occurrence rate of the aforementioned error pattern was checked.A possible error accumulation or even a systematic error, which leads to a corrective action of the installed base, could not be determined by the investigation.
 
Additional Manufacturer Narrative
7/19/2024: supplemental mdr 2 was submitted to the fda for correction of the primary udi number in section d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE FLOOR
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemensstrasse 1 or
rittigfeld 1
forchheim 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHINEERS AG
siemensstrasse 1 or
rittigfeld 1
forchheim 91301
GM   91301
Manufacturer Contact
anastasia sokolova
40 liberty blvd.
malvern, PA 19355
4843234197
MDR Report Key18152317
Report Number3004977335-2023-00151
Device Sequence Number6467318
Product Code OWB
UDI-Device Identifier04056869010045
UDI-Public(01)04056869010045(21)137064
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K181407
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10094135
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/31/2023
Supplement Date Received by Manufacturer12/18/2023
07/19/2024
Initial Report FDA Received Date11/16/2023
Supplement Report FDA Received Date12/18/2023
07/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-