| Catalog Number |
1588010000 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: cib, jamie, asr, losing picture.[update - replacement device used to complete the procedure.Full loss of image, but the duration cannot be confirmed.] the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause/s could be a bad 1588 camera, potential electrical failure on digital board, bad cables,and/or loose cable connections.Advise to calibrate unit.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|