Device evaluation summary: visual analysis of the returned sample revealed that the sx mpp gel 375cc breast implant have a tear on the posterior view within an area of siltex cracking, measuring approximately 0.1 cm.The evaluation determined that the possible cause of the rupture found is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.Mentor concluded that the capsular contracture in the patient´s breast was the result of the body´s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.A seroma is a build-up of fluid around the implant.Symptoms from a seroma may include swelling, pain, and bruising.A seroma may occur soon after surgery, or any time later on, which would be referred to as a late seroma.While the body may absorb seromas, some may require surgery for drainage.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture and late onset seroma.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
|
It was reported that a 33-year-old caucasian female patient underwent a primary breast augmentation with two 375cc mentor memorygel breast implants and experienced bilateral capsular contracture (baker grade 3) post-operatively, which was diagnosed with an office visit.It was also reported that the patient developed late onset seroma in the right breast.Seroma fluid was collected and sent to pathology/cytology for testing.The final cytology report is negative for malignant cells.As a result, the patient underwent bilateral explantation on (b)(6) 2023.This report is for the right side device.
|