| Catalog Number |
420915 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that post op x-rays indicate patient outcome resulted in 27 degrees of flexion.Surgeon is dissatisfied with the result of the planning.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = per sales rep, post op x-rays indicate patient outcome resulted in 27 degrees of flexion.Surgeon is dissatisfied with the result of the planning.The product was not returned to depuy synthes, however photos were provided for review.The x-ray investigation revealed that the provided photo does not represent the reported device.The overall complaint cannot be confirmed.A product development investigation was performed and it was concluded that the segmentation is designed within trumatch specifications and no issues were identified.A manufacturing record evaluation was performed for the finished device product description: trumatch ct cut guide kit l product code: 420915 lot number: 00063281 and no non-conformances or manufacturing irregularities were identified.The overall complaint was unconfirmed as the observed condition of the trumatch ct cut guide kit l would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot = a manufacturing record evaluation was performed for the finished device product description: trumatch ct cut guide kit l product code: 420915 lot number: 00063281 and no non-conformances or manufacturing irregularities were identified.Device history review = a manufacturing record evaluation was performed for the finished device product description: trumatch ct cut guide kit l product code: 420915 lot number: 00063281 and no non-conformances or manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|