| Catalog Number |
36010 |
| Medical Device Problem Code |
Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a spectrum pump alarmed "a34831, h0629 corrective action".This occurred during an unspecified process step.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for evaluation.Functional testing was performed and did not identify any issues related to the customer reported condition.The device functioned as intended when installed in a known working pump, however, evaluation determined that the wbm will be scrapped.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.The device malfunction would not lead to a death or serious injury if the malfunction were to recur.
|
| |
|