| Catalog Number |
64811100 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Codes |
Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
|
| Date of Event |
10/25/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's left knee was revised due to the stem breaking at the junction with the femoral component of a hinged knee construct.Revising surgeon reported he feels the femoral side should have been a distal femoral replacement.The patient's mrs/ mrh knee components (except for the proximal tibial component and stem) were revised.Rep provided all information that was available from the hospital and surgeon.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding crack/fracture involving a mrh femoral component was reported.The event was confirmed via visual inspection of the provided photographs of the device and clinician review.Method & results: product evaluation and results: the reported device was not returned however photographs were provided for review.Visual inspection of the provided photographs of the devices indicate that the devices have been recently explanted and are covered in blood/tissue and bone cement.A portion of the threaded junction of the femoral component appears to have fractured off.Clinician review: a review of medical records with a clinical consultant indicated "the ap and lateral x-rays show that hinged femoral component has fractured at its connection to the stem.Stem breaking at the junction with the femoral component of a hinged knee construct is confirmed.No documentation was available to confirm that a revision was completed.The root cause of the stem breakage cannot be determined from the limited documentation provided.Should additional information become available i would be happy to further this assessment." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to fracture at the stem/femoral component interface.The reported device was not returned however photographs were provided for review.Visual inspection of the provided photographs of the devices indicate that the devices have been recently explanted and are covered in blood/tissue and bone cement.A portion of the threaded junction of the femoral component appears to have fractured off.A review of medical records with a clinical consultant indicated "the ap and lateral x-rays show that hinged femoral component has fractured at its connection to the stem.Stem breaking at the junction with the femoral component of a hinged knee construct is confirmed.No documentation was available to confirm that a revision was completed.The root cause of the stem breakage cannot be determined from the limited documentation provided.Should additional information become available i would be happy to further this assessment." no further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's left knee was revised due to the stem breaking at the junction with the femoral component of a hinged knee construct.Revising surgeon reported he feels the femoral side should have been a distal femoral replacement.The patient's mrs/ mrh knee components (except for the proximal tibial component and stem) were revised.Rep provided all information that was available from the hospital and surgeon.
|
| |
|