| Model Number |
1116985 |
| Medical Device Problem Codes |
Smoking (1585); Sparking (2595)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
N/an/a device not yet returned.
|
| |
|
Event or Problem Description
|
|
13-nov-2023: the technical consultant emailed technical services (ts).The customer reported spark coming from the cartridge holder/compartment of their afinion 2, right after there was a smell of burnt.We have powered off the machine.The customer also provided the following information: serial number: (b)(6). software version: w 21.14 no harm to user circuit breaker did not trip.
|
| |
|
Additional Manufacturer Narrative
|
|
As of 11.02.2025 the instrument has not been received by the legal manufacturer for further investigation.Customer reported some sort of spark coming from the cartridge holder/compartment of the afinion 2 serial number (b)(6) (sw: 21.14).Right after there was a smell of burnt.The analyzer was powered off after the event.No harm reported.The following steps were taken by the manufacturer: review of (b)(6) history records; similar complaint trend analysis; product risk review; review of non-conformance data.The complaint details were reviewed along with internal records.There were no adverse patient effects and no clinically significant delay in the procedure reported in the complaint.Our investigation determined that there is no indication that the reported issue is related to product results, inadequate instructions for use or inaccurate labeling.Our investigation did determine that the spark allegations was a led light flash from inside the cartridge chamber, which is considered normal behavior when removing the power cable from the instrument.It is not possible for the inside of the instrument to produce a spark.The event of the burnt smell was determined to be caused by an analyzer failure.Furthermore, it should be noted that the analyzer might be impaired by the following: condensing humidity and water, heat and large temperature variations, direct sunlight, vibrations (e.G.From centrifuges and dishwashers), electromagnetic radiation, movement of the analyzer during processing of a test cartridge.
|
| |
|
Event or Problem Description
|
|
13-nov-2023: the technical consultant emailed technical services (ts).The customer reported spark coming from the cartridge holder/compartment of their afinion 2, right after there was a smell of burnt.We have powered off the machine.The customer also provided the following information: serial number: (b)(6).Software version: w 21.14.No harm to user.Circuit breaker did not trip.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspection of the instrument was performed and concluded 29.07.2025.The overall inspection passed, however the instrument will be scrapped according procedure due to complainant observation of sparks and electrostatic.
|
| |
|
Event or Problem Description
|
|
13-nov-2023: the technical consultant emailed technical services (ts).The customer reported spark coming from the cartridge holder/compartment of their afinion 2.Right after there was a smell of burnt.The customer powered off the machine.The customer also provided the following information: serial number: (b)(6).Software version: sw 21.14.No harm to user.Circuit breaker did not trip.
|
| |
|