• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS TINA-QUANT HEMOGLOBIN A1CDX GEN. 3; GLYCOSYLATED HEMOGLOBIN ASSAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS TINA-QUANT HEMOGLOBIN A1CDX GEN. 3; GLYCOSYLATED HEMOGLOBIN ASSAY Back to Search Results
Catalog Number 08056668190
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The serial number of the c 503 analyzer is (b)(6).The sample probe assembly was changed and adjusted.The investigation is ongoing.
 
Event or Problem Description
The initial reporter stated they received discrepant results for three patient samples tested with tina-quant hemoglobin a1cdx gen.3 (hba1c) on a cobas pro c 503 analytical unit.The first sample initially resulted in an hba1c value of 7.81 % and it repeated as 5.45 %.The second sample initially resulted in an hba1c value of 7.34 % and it repeated as 5.77 %.The third sample initially resulted in an hba1c value of 5.95 % and it repeated as 5.40 %.
 
Additional Manufacturer Narrative
The issue was resolved after the exchange and adjustment of a sample probe.No further issues have occurred.The investigation determined the exchange and adjustment of the probe resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TINA-QUANT HEMOGLOBIN A1CDX GEN. 3
Common Device Name
GLYCOSYLATED HEMOGLOBIN ASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18158758
Report Number1823260-2023-03646
Device Sequence Number12762279
Product Code LCP
UDI-Device Identifier07613336120927
UDI-Public07613336120927
Combination Product (Y/N)Y
Initial Reporter CountryGR
PMA/510(K) Number
K072714
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number08056668190
Device Lot Number715056
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2023
Supplement Date Received by Manufacturer11/17/2023
Initial Report FDA Received Date11/17/2023
Supplement Report FDA Received Date11/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-