| Catalog Number |
08056668190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/26/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The serial number of the c 503 analyzer is (b)(6).The sample probe assembly was changed and adjusted.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for three patient samples tested with tina-quant hemoglobin a1cdx gen.3 (hba1c) on a cobas pro c 503 analytical unit.The first sample initially resulted in an hba1c value of 7.81 % and it repeated as 5.45 %.The second sample initially resulted in an hba1c value of 7.34 % and it repeated as 5.77 %.The third sample initially resulted in an hba1c value of 5.95 % and it repeated as 5.40 %.
|
| |
|
Additional Manufacturer Narrative
|
|
The issue was resolved after the exchange and adjustment of a sample probe.No further issues have occurred.The investigation determined the exchange and adjustment of the probe resolved the issue.
|
| |
|