• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® FOLEY TRAY WITH LUBRI-SIL® ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® FOLEY TRAY WITH LUBRI-SIL® ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 947316
Medical Device Problem Codes Gel Leak (1267); Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the inside this product comes an accessory, called urinary catheter lubrication gel.This accessory was chemically changed, that is, the client indicates that it came dry inside the packaging, for this reason it could not be used, that is, this lubricant was used so that the urinary catheter passes through the urinary meatus without resistance and more soft.It was stated that no impact on the patient.
 
Event or Problem Description
It was reported that the inside this product comes an accessory, called urinary catheter lubrication gel.This accessory was chemically changed, that is, the client indicates that it came dry inside the packaging, for this reason it could not be used, that is, this lubricant was used so that the urinary catheter passes through the urinary meatus without resistance and more soft.It was stated that no impact on the patient.
 
Additional Manufacturer Narrative
The reported event is confirmed cause unknown.Visual evaluation noted received 3 photo samples.First photo sample shows top overview of product packaging with damaged gel syringe next to product packaging.Second photo sample shows top view of damaged syringe and product lot number and expiration date.Third photo sample shows close up of damaged syringe with gel remnants on the side of the syringe.Based on the sample received product does not meet specifications which states "missing, damaged, leaking, ripped, cracked and incomplete components are not allowed.".The exact root cause could not be identified.Although a root cause could not be definitively identified, based on the risk documentation review, a potential root cause for this type of failure could be inappropriate package design.However, there was insufficient information to confirm this potential root cause.A dhr review did not show any problems or conditions that would have contributed to the reported event.The labelling review is not required as labelling would not have prevented the reported event.Section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® FOLEY TRAY WITH LUBRI-SIL® ALL-SILICONE FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18159736
Report Number1018233-2023-08235
Device Sequence Number8045753
Product Code EZL
UDI-Device Identifier00801741024931
UDI-Public(01)00801741024931
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K040504
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Catalogue Number947316
Device Lot NumberNGGU3904
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/24/2023
Supplement Date Received by Manufacturer03/28/2024
Initial Report FDA Received Date11/17/2023
Supplement Report FDA Received Date03/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-