• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RENEGADE HI-FLO FATHOM SYSTEM; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RENEGADE HI-FLO FATHOM SYSTEM; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 29462
Medical Device Problem Codes Fracture (1260); Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the mid shaft of the device was fractured.A renegade hi-flo fathom system was selected for use.During advancing, the physician felt resistance when manipulating the microcatheter so the physician tried to retract it, but it would not retract more than about 10 cm.When the base catheter and the microcatheter were removed from the patient, it was noted that the mid shaft of the microcatheter was fractured.The detached portion of the device was removed from the body and no device fragments were left inside the body.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.The device was visually and microscopically inspected for damage.The renegade device showed 3 areas of damage.One fracture located 26.9cm, stretching of the shaft located 30cm from the hub and 1 stretched and fractured area located 34.3cm from the hub.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The complaint was confirmed for advancing issues and shaft damage.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that the mid shaft of the device was fractured.A renegade hi-flo fathom system was selected for use.During advancing, the physician felt resistance when manipulating the microcatheter so the physician tried to retract it, but it would not retract more than about 10 cm.When the base catheter and the microcatheter were removed from the patient, it was noted that the mid shaft of the microcatheter was fractured.The detached portion of the device was removed from the body and no device fragments were left inside the body.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEGADE HI-FLO FATHOM SYSTEM
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18161738
Report Number2124215-2023-63503
Device Sequence Number7275137
Product Code KRA
UDI-Device Identifier08714729790860
UDI-Public08714729790860
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K140329
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/24/2026
Device Model Number29462
Device Catalogue Number29462
Device Lot Number0031696509
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/02/2023
Supplement Date Received by Manufacturer12/21/2023
Initial Report FDA Received Date11/17/2023
Supplement Report FDA Received Date01/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-