| Model Number |
24674 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: mustang 10.0 x 60, 75cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A balloon longitudinal tear was identified beginning approximately 25mm distal of the proximal markerband and extending approximately 11mm distally across the balloon material.The rated burst pressure for this device is 14 atmospheres.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual and tactile examination found no damage to the tip of the device.A visual examination observed no issues with the markerbands of the device.Both markerbands were undamaged in the correct position on the device.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 15-nov-2023.It was reported that advancing difficulties were encountered.Vascular access was obtained via contralateral approach.The 100% stenosed target lesion was located in the severely tortuous common iliac artery.A 10.0 x 60, 75cm mustang balloon catheter was advanced for dilation.However, during the procedure, the device did not move in the opposite direction during the contralateral approach process.The procedure was completed with another of the same device.No patient complications were reported.However, returned device analysis revealed balloon longitudinal tear.
|
| |
|