| Catalog Number |
730M0024 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that intra-op during the derotation procedure, a vital rocket reducer was damaged.One of the two blocking pins was pulled off.The procedure was completed without harm to patient using other rocket reducers.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed one of the pins connecting the arm to the body has disassembled from the device.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review: per dhr review, the part was likely conforming when it left highridge control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that intra-op during the derotation procedure, a vital rocket reducer was damaged.One of the two blocking pins was pulled off.The procedure was completed without harm to patient using other rocket reducers.
|
| |
|