• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. SHORT ROCKET REDUCER; VITALITY SPINAL FIXATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. SHORT ROCKET REDUCER; VITALITY SPINAL FIXATION SYSTEM Back to Search Results
Catalog Number 730M0024
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/25/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event or Problem Description
It was reported that intra-op during the derotation procedure, a vital rocket reducer was damaged.One of the two blocking pins was pulled off.The procedure was completed without harm to patient using other rocket reducers.
 
Additional Manufacturer Narrative
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Additional information in h6: component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed one of the pins connecting the arm to the body has disassembled from the device.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.Dhr review: per dhr review, the part was likely conforming when it left highridge control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that intra-op during the derotation procedure, a vital rocket reducer was damaged.One of the two blocking pins was pulled off.The procedure was completed without harm to patient using other rocket reducers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHORT ROCKET REDUCER
Common Device Name
VITALITY SPINAL FIXATION SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
sabrina abla
10225 westmoor dr.
westminster, CO 80021
7206965158
MDR Report Key18162751
Report Number3012447612-2023-00354
Device Sequence Number7275176
Product Code NKB
UDI-Device Identifier00887868244780
UDI-Public(01)00887868244780(10)MC4395804
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K183550
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number730M0024
Device Lot NumberMC4395804
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer11/02/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date11/17/2023
Supplement Report FDA Received Date06/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/21/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-