• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 1012530-60
Medical Device Problem Codes Premature Activation (1484); Difficult to Remove (1528); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3, b6: dates are estimated.Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event or Problem Description
It was reported that during the procedure, the guide wire was in the target lesion and the prepared absolute pro self-expanding stent system (sess) was being advanced over the guide wire, when the stent became exposed [started to deploy] prior to entering the introducer sheath.The sess was then attempted to be removed from the guide wire, however resistance was met and the outer sheath of the sess was noted as bunched up.They decided to deploy the stent on the guide wire in order to remove it.A new absolute pro stent was used successfully for the procedure.There were no reported adverse patient effects and no clinically significant delays in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported deformation due to compressive stress was able to be confirmed.The reported premature activation and the reported difficult to remove were unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during advancement onto the guide wire inadvertent mishandling resulted in the reported bunched outer sheath/noted wrinkled sheath, the noted kinked/bunched inner member and the noted smashed/torn tip; thus compromising the device resulting in the reported premature activation and the reported difficult to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18162966
Report Number2024168-2023-12844
Device Sequence Number17942102
Product Code NIP
UDI-Device Identifier08717648175718
UDI-Public08717648175718
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue Number1012530-60
Device Lot Number3060961
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2023
Supplement Date Received by Manufacturer01/03/2024
Initial Report FDA Received Date11/17/2023
Supplement Report FDA Received Date01/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-