| Catalog Number |
3505-7540 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, four closure tops were stripped and two screws were disassembled.New closure tops and screws were used to complete the case.There was a 30-minute delay without any impact on the patient.This is report six of six for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, four closure tops were stripped and two screws were disassembled.New closure tops and screws were used to complete the case.There was a 30-minute delay without any impact on the patient.This is report six of six for this event.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, four closure tops were stripped and two screws were disassembled.New closure tops and screws were used to complete the case.There was a 30-minute delay without any impact on the patient.This is report six of six for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Component, investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tulip has disassembled from the screw shank.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to off-axis forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference reports 3012447612-2023-00356 through 3012447612-2023-00361.
|
| |
|