Catalog Number 1012270-20
Medical Device Problem Codes Break (1069); Stretched (1601); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.The nc trek device referenced in b5 is filed under separate a medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat calcified lesions in the middle and distal third right coronary artery (rca).The hi-torque 014 bmw guidewire (gw) was used in the procedure; however, the gw would not allow the advancement of a 2.0 mini trek balloon dilatation catheter (bdc) and a 3.25x20mm nc trek bdc.When attempting to advance each bdc, resistance was met with the gw and the shaft became stretched and broke on each bdc.The gw was exchanged for a hi-torque 014 bhw gw, which was also noted to have difficulty when advancing other devices.Two drug-eluting stents were implanted to complete the procedure.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported difficulties.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18164098
Report Number2024168-2023-12852
Device Sequence Number15284233
Product Code LOX
UDI-Device Identifier08717648138164
UDI-Public08717648138164
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Catalogue Number1012270-20
Device Lot Number20902G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/26/2023
Supplement Date Received by Manufacturer11/22/2023
Initial Report FDA Received Date11/17/2023
Supplement Report FDA Received Date12/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HI-TORQUE 014 BMW HYDRO 3CM J GUIDEWIRE
Patient Age81 YR
Patient SexUnknown
-
-