Catalog Number 71421533
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a tkr surgery, the gii cr deep flex isrt s1-2 13m couldn't lock tightly.The procedure was resumed, after a significant delay, with a s+n back-up device.The patient required additional anesthesia.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
H10: internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device reveals scratches and gouges in the surface of the device.The dimensional evaluation revealed that the device has damage that appears to be from attempted use and has the potential to cause an event during attempted mating to the tibial base as described in the complaint.The damage/deformation of several features would not allow for accurate measurement.All measurable critical features that could be measured were within specification.The dimensional inspection of the returned device did not indicate the product was defective at the time of shipment from the manufacturing site.This device is subjected to multiple visual and dimensional inspection during processing, and this type of damage has a high likelihood of detection during these inspections.Based on the evidence provided, the unsatisfactory experience could be confirmed.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.Factors that could contribute to the reported event include size selected or insertion technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII CR DEEP FLEX ISRT S1-2 13M
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18164676
Report Number1020279-2023-02258
Device Sequence Number6466464
Product Code JWH
UDI-Device Identifier03596010520364
UDI-Public03596010520364
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K041825
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71421533
Device Lot Number19HM00230
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2023
Supplement Date Received by Manufacturer12/28/2023
Initial Report FDA Received Date11/17/2023
Supplement Report FDA Received Date01/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/01/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age65 YR
Patient SexFemale
-
-