| Catalog Number |
72200872 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/1901
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.A functional evaluation revealed a blade stall error and overheating of the housing.Further evaluation revealed the drive fork was stiff when rotated.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with a mechanical component failure.Factors that could have contributed to the failure include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This can be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.No containment or corrective actions are recommended at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that during set up of a procedure, the motor drive unit was not responding to foot or hand control in port a or b.Back up device was available.No surgical delay nor further complications were reported.Per the results of the investigation, the functional evaluation revealed an overheating of the housing.
|
| |
|
Search Alerts/Recalls
|