| Model Number |
08946353190 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cobas e411 rack serial number was (b)(6).The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable testosterone g2 elecsys results from the cobas e411 rack analyzer.Patient 1 initial result was >1500 ng/dl and the repeat results were 69.83 ng/dl and 46.45 ng/dl.The repeat results on another analyzer were 28.62 ng/dl and 33.47 ng/dl.Patient 2 initial result was 177.5 ng/dl and the repeat results were 46.47 ng/dl and 109.8 ng/dl.The repeat results on another analyzer were 38.24 ng/dl and 39.83 ng/dl.The results from the other analyzer were believed correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer took samples of dust in the laboratory and also swabbed the area around the instrument, as well as sample/reagent probe.The samples were tested at a nearby facility and each one resulted as greater than the measuring range for testosterone.The field service engineer determined the cause of the event was the testosterone contamination, likely coming from the hyase product used in the ivf lab on site.He replaced the sample/reagent probe and pressure sensor tubing.He performed sample/reagent probe, sipper, and gripper adjustments.He replaced the measuring cell and syringe assembly and ran performance testing that passed.As the qc recovery was within the acceptable range, there was no indication of a performance issue with the reagent.
|
| |
|
Additional Manufacturer Narrative
|
|
**udi related data quality updates only** d3 is the initial distributor in the us.
|
| |
|