MAUDE Adverse Event Report: MICROVENTION, INC. SOFIA FLOW PLUS 6F; CATHETER, THROMBUS RETRIEVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROVENTION, INC. SOFIA FLOW PLUS 6F; CATHETER, THROMBUS RETRIEVER Back to Search Results
Model Number UNKNOWN
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Intracranial Hemorrhage (1891)
Date of Event 03/05/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A search for non-conformances associated with the reported part and lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was discarded and not available for return to the manufacturer for analysis; however, medical notes were provided and are being reviewed part of ongoing investigation.A supplemental report will be submitted once investigation is completed.The instructions for use (ifu) identifies intracerebral/intracranial hemorrhage as a potential complication associated with the use of the device.
 
Event or Problem Description
As reported through the sofast clinical study, the next day following mechanical thrombectomy procedure that used the sofia aspiration catheter, the patient experienced a subarachnoid hemorrhage noted on ct of the head and described as mild, asymptotic, no change in nihss.Th patient was administered iv heparin, intra-arterial tpa and resolved with sequelae.The event is assessed as possibly related to the sofia device, definitely related to the mechanical thrombectomy procedure, and possibly related to the study disease and concurrent condition/treatment.
 
Additional Manufacturer Narrative
Summary of investigation: visual analysis: a visual inspection of the device captured in this file could not be performed as a physical device was not returned for evaluation.Investigation findings: without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.Based on a review of the device¿s risk documentation, the reported event did not indicate there were any potential or new manufacturing, design, quality, or other systemic issues, or non-conformances.The complaint code is monitored through the trending process; corrective action is determined, as needed, through this process.Investigations of historic complaint files with similar complaint category coding are recorded in the complaint handling system; without the ability to perform and analysis of the device, this investigation cannot identify with certainty any potential root causes.Batch review: a search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.Complaint system review: there are no similar complaints based on the complaint category regarding this batch number from the last two years recorded in the complaint system at the time of this investigation.Ifu review (additional information can be found in the ifu): potential complications potential complications include but are not limited to: vasospasm, hematoma at the site of entry, embolism, ischemia, intracerebral/intracranial hemorrhage, pseudo aneurysm, seizure, stroke, infection, vessel dissection, thrombus formation, and death.Precautions exercise care in handling the sofia flow plus aspiration catheter to reduce the chance of accidental damage.Use caution when manipulating the sofia flow plus aspiration catheter in tortuous vasculature to avoid damage.Avoid advancing or withdrawal against resistance until the cause of resistance is determined.The presence of calcification, irregularities, or other devices may damage the sofia flow plus aspiration catheter and potentially affect its insertion or removal.Maintain perfusion of heparinized saline for inner lumen of the sofia flow plus aspiration catheter to prevent thrombus formation.Do not use automated high-pressure contrast injection equipment with the sofia flow plus aspiration catheter because it may damage the catheter.The catheter has hydrophilic coating over the distal 60 cm.Make sure to hydrate the coating before use with heparinized saline.Once the catheter is hydrated, do not allow it to dry.Delivery of the sofia flow plus aspiration catheter 6.Navigation through the vasculature b.Insert the guidewire and / or microcatheter into the sofia flow plus aspiration catheter and advance the guidewire / microcatheter until the guidewire / microcatheter and the sofia flow plus aspiration catheter are aligned at the distal end.C.Using the introducer sheath provided in the package, carefully insert the sofia flow plus aspiration catheter and the guidewire through a hemostatic valve of the femoral sheath.Warning: do not over-tighten the hemostatic valve on the sheath or guide catheter through which the sofia flow plus aspiration catheter is inserted.Over-tightening may result in damage to the sofia flow plus aspiration catheter.E.Under fluoroscopic guidance, advance or withdraw the sofia flow plus aspiration catheter over the guidewire and / or microcatheter until the desired position is attained.Select vessels by slowly torqueing the sofia flow plus aspiration catheter if necessary.Warning: do not advance or withdraw the device when excessive resistance is met until the cause of resistance is determined.Warning: torqueing the sofia flow plus aspiration catheter excessively while kinked may damage the device resulting in separation of the device.Withdraw the entire device (the device, microcatheter, and guidewire) if the device is severely kinked.Aspiration through the sofia flow plus aspiration catheter 8.Under fluoroscopic guidance, position the distal tip of the sofia flow plus aspiration catheter at the desired vessel location.Warning: do not advance or withdraw the device when excessive resistance is met until the cause of resistance is determined.Warning: torqueing the sofia flow plus aspiration catheter excessively while kinked may damage the catheter resulting in separation of the catheter.Withdraw the catheter system, including the sofia flow plus aspiration catheter, microcatheter, and guidewire if the sofia flow plus aspiration catheter is severely kinked.9.Attach aspiration tubing to the aspiration pump and turn on the pump (refer to the directions for use of the aspiration tubing and aspiration pump manual).Confirm that the aspiration gauge reads -20 inhg.Ensure the stopcock on the aspiration tubing is in the closed position.13.To begin aspiration, turn the aspiration tubing stopcock to the open position and check to see if blood, thrombus, or embolus are aspirated through the system.14.If after 10 seconds blood is still observed flowing through the system, stop aspiration.To stop aspiration, turn aspiration tubing stopcock to the closed position.Carefully reposition the distal tip of the sofia flow plus aspiration catheter to engage the thrombus and resume aspiration.15.If flow is restricted or absent, maintain aspiration to make sure any thrombus or embolus is fully engaged with the distal tip of the sofia flow plus aspiration catheter.With the thrombus or embolus fully engaged, slowly pull back the sofia flow plus aspiration catheter and completely withdraw out of the patient.Warning: excessive aspiration with the distal tip of the sofia flow plus aspiration catheter engaged with vessel wall may cause vessel injury.16.With the sofia flow plus aspiration catheter outside of the patient body, flush the catheter to clear the catheter of any thromboembolic material that may be inside.Warning: do not attempt to clear inner lumen of the sofia flow plus aspiration catheter by infusion while keeping the device in the patient body.Remove the sofia flow plus aspiration catheter from the patient body before attempting to clear the lumen.17.If repeated access to the vasculature with the same device is desired, flush and clean the inner lumen of the device by infusion.Inspect the device for any damage.Reintroduce the sofia flow plus aspiration catheter into the body and follow steps 6 & 7 in the ¿delivery of the sofia flow plus aspiration catheter¿ section to navigate the catheter to the target site.Warning: do not use the device if any damage or irregularities are observed.Investigation conclusion: a medical note review was performed.Review of data indicates that the medical note data is dated (b)(6) 2023.As part of the sofast clinical study, a mechanical thrombectomy procedure was performed with the utilization of a sofia 6f aspiration catheter device on (b)(6) 2023.On (b)(6) 2023, reported postoperative day one, the patient experienced a reported subarachnoid hemorrhage throughout the left frontal parietal region and extending into the left basal cistern which was confirmed by ct imaging.Thrombectomy procedure was reported as completed successfully with resolution of sah with patient discharged to home on (b)(6) 2023 with physical therapy and occupational therapy prescribed.Based on the review, the sah is possibility related to the use of the sofia device which cannot be ruled out, and the sah is identified as related to the performance of the thrombectomy procedure.However, no device malfunction was reported and /or no device malfunction was reported as associated with or contributing to the reported sah.Without the return and physical evaluation of the device, the investigation cannot definitively determine if a condition existed that would have caused or contributed to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOFIA FLOW PLUS 6F
Common Device Name
CATHETER, THROMBUS RETRIEVER
Manufacturer (Section D)
MICROVENTION, INC.
35 enterprise
aliso viejo CA 92656
Manufacturer Contact
terrence callahan
35 enterprise
aliso viejo, CA 92656
7142478000
MDR Report Key18166010
Report Number2032493-2023-01058
Device Sequence Number20636808
Product Code NRY
UDI-Device Identifier00842429108482
UDI-Public(01)00842429108482(11)230110(17)251231(10)0000307038
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K173200
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberUNKNOWN
Device Lot Number0000307038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2023
Supplement Date Received by Manufacturer12/15/2023
Initial Report FDA Received Date11/17/2023
Supplement Report FDA Received Date12/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-