| Model Number |
N161 |
| Medical Device Problem Code |
Failure to Capture (1081)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the left ventricular (lv) lead of this cardiac resynchronization therapy defibrillator (crt-d) system exhibited loss of capture (loc).Technical services (ts) reviewed troubleshooting options.This crt-d system remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
It was reported that the left ventricular (lv) lead of this cardiac resynchronization therapy defibrillator (crt-d) system exhibited loss of capture (loc).Technical services (ts) reviewed troubleshooting options.This crt-d system remains in service.No adverse patient effects were reported.
|
| |
|