• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INCEPTA CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number N161
Medical Device Problem Code Failure to Capture (1081)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the left ventricular (lv) lead of this cardiac resynchronization therapy defibrillator (crt-d) system exhibited loss of capture (loc).Technical services (ts) reviewed troubleshooting options.This crt-d system remains in service.No adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Event or Problem Description
It was reported that the left ventricular (lv) lead of this cardiac resynchronization therapy defibrillator (crt-d) system exhibited loss of capture (loc).Technical services (ts) reviewed troubleshooting options.This crt-d system remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INCEPTA CRT-D
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18166423
Report Number2124215-2023-65159
Device Sequence Number12427863
Product Code LWP
UDI-Device Identifier00802526481000
UDI-Public00802526481000
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/14/2016
Device Model NumberN161
Device Catalogue NumberN161
Device Lot Number501657
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/08/2023
Supplement Date Received by Manufacturer07/24/2025
Initial Report FDA Received Date11/17/2023
Supplement Report FDA Received Date08/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexMale
-
-