| Model Number |
401771 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Arrhythmia (1721)
|
| Date of Event |
10/21/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During treatment for third-degree atrioventricular block, bradycardia occurred requiring medication administration to stabilize the patient.A temporary pacemaker was placed via the right femoral artery.Later that morning, the ecg monitor alarmed indicating bradycardia.Upon inspection, the connection line of the temporary pacemaker catheter was noted to be broken.Medication was administered to treat the bradycardia, the patient stabilized and the catheter was removed.
|
| |
|
Event or Problem Description
|
|
This report is to include the name of the physician involved in the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
An image was submitted for evaluation to product performance engineering; no device components were returned.The ¿(+)¿ proximal leg braid wire appeared to have fractured near the bifurcation area, resulting in wire exposure and separation of the leg wire tubing from the connector pin assembly.Visual inspection was based solely upon a review of the photograph provided.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported bradycardia device damage could not be conclusively determined.
|
| |
|