| Model Number |
11500A25 |
| Medical Device Problem Codes |
Calcified (1077); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
|
| Date of Event |
10/31/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported a patient with a 25mm 11500a aortic valve implanted four (4) years, one (1) month, is being evaluated for valve-in-valve procedure due to severe aortic insufficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to edwards for evaluation as it remains implanted.Attempts to retrieve additional information is in process.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.
|
| |
|
Additional Manufacturer Narrative
|
|
Added information to b5, h6.
|
| |
|
Event or Problem Description
|
|
It was reported a patient with a 25mm 11500a aortic valve implanted four (4) years, one (1) month, underwent valve-in-valve procedure due to severe aortic insufficiency.Tavr was successfully performed with a 26mm 9755rsl transcatheter valve.
|
| |
|
Event or Problem Description
|
|
It was reported a patient with a 25mm 11500a aortic valve implanted four (4) years, one (1) month, underwent valve-in-valve procedure due to severe aortic insufficiency and moderate calcification.Patient presented with congestive heart failure.Tavr was successfully performed with a 26mm 9755rsl transcatheter valve.Patient tolerated the procedure well and was transferred to ccu in stable condition.Patient was discharged home on pod # 1.
|
| |
|
Additional Manufacturer Narrative
|
|
Added information to b5, b6, b7, h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Calcific degeneration is a common cause of bioprosthetic heart valve failures.Many factors contribute to the onset and propagation of calcification.These include patient factors (age, disease state, pharmacological intervention, etc.), mechanical stress related to the valves hemodynamic performance, and glutaraldehyde fixation of tissue.Of these, the fixation process is a relatively minor contributor to calcification for edwards tissue valves due to anti-calcification treatments during manufacturing.Though numerous studies have been conducted on preventive calcification strategies in bioprosthetic heart valves, the causes of calcification are not fully understood and there are still no mechanisms or medical therapies which fully prevent bioprostheses from calcifying.Calcification is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.The most likely cause is patient factors, including a history of dyslipidemia and non-obstructive cad.
|
| |
|