| Model Number |
N/A |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Fall (1848); Joint Dislocation (2374)
|
| Date of Event |
10/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient underwent a revision procedure two years post implantation due to dislocation.Patient had a fall at some point and then had an episode of dislocation.The hip was attempted to be relocated in the emergency department - surgeon feels at this point the poly 40mm head was pulled off the 28mm bearing and then was sitting loose in the joint space.The hip was revised - g7 cup stayed in and the dm liner was revised to a freedom constrained liner and head.It was noted during revision it looked like the repair of the external rotators of the hip had failed some time ago.There is no additional information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: 00801802803 femoral head sterile product do not resterilize 12/14 taper lot number is 64730408.G2: foreign: australia.The device will not be returned for analysis as it¿s location is not known; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified the explanted bearing covered in bio-debris, with some scratching noted to the outer spherical surface.No further evaluation can be made from the provided picture.Review of the device history records identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: right total hip arthroplasty with evidence of leg length discrepancy on the initial image and posterior and superior dislocation of the right femoral head on the second image.A definitive root cause cannot be determined.It was noted the patient fell leading to the dislocation, however the reason for the fall is unknown if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
There is no update to the prior event description provided.
|
| |
|