ZIMMER BIOMET, INC. ACT ARTIC E1 HIP BRG 28X40MM; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Fall (1848); Joint Dislocation (2374)
Date of Event 10/23/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a revision procedure two years post implantation due to dislocation.Patient had a fall at some point and then had an episode of dislocation.The hip was attempted to be relocated in the emergency department - surgeon feels at this point the poly 40mm head was pulled off the 28mm bearing and then was sitting loose in the joint space.The hip was revised - g7 cup stayed in and the dm liner was revised to a freedom constrained liner and head.It was noted during revision it looked like the repair of the external rotators of the hip had failed some time ago.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).D10: 00801802803 femoral head sterile product do not resterilize 12/14 taper lot number is 64730408.G2: foreign: australia.The device will not be returned for analysis as it¿s location is not known; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified the explanted bearing covered in bio-debris, with some scratching noted to the outer spherical surface.No further evaluation can be made from the provided picture.Review of the device history records identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: right total hip arthroplasty with evidence of leg length discrepancy on the initial image and posterior and superior dislocation of the right femoral head on the second image.A definitive root cause cannot be determined.It was noted the patient fell leading to the dislocation, however the reason for the fall is unknown if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACT ARTIC E1 HIP BRG 28X40MM
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18168807
Report Number0001825034-2023-02665
Device Sequence Number5541085
Product Code LPH
UDI-Device Identifier00880304485037
UDI-Public(01)00880304485037(17)251002(10)789500
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K161190
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/02/2025
Device Model NumberN/A
Device Catalogue NumberEP-200146
Device Lot Number789500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2023
Supplement Date Received by Manufacturer02/28/2024
Initial Report FDA Received Date11/20/2023
Supplement Report FDA Received Date03/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;