| Model Number |
N/A |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported, that the customer had opened the original sealed box and went to open the screw package, but noticed the screw appears to be melted to the inner packaging.No further information at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Reported event was confirmed as visual evaluation of the returned product found the pouch exhibits damage that is visually consistent with thermal damage.Outer pouch visually appears to be adhered to the inner pouch.Sterility has been breached.This complaint has been confirmed by evaluation of the returned product.Device history record was reviewed and no discrepancies relevant to the reported event were found.The condition of the device when it left zimmer biomet is considered non-conforming to specification.The root cause of the reported event can be attributed to packaging operator not following the work instructions provided.Corrective actions have been initiated.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|