| Model Number |
B31040 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that tested the lead intra-op the impedances were all high.The lead cabled was swapped and the impedances resolved.It was tried 3 times before switching the cable there was no change.Actions taken were swapped the cable after reconnecting 3 times, pushing the stylet in, and resisting the cable with the lead test cable jaws closed.The issue was resolved.No symptoms reported.Additional information was received from the manufacturer representative (rep).The cause was not determined.The stylet was pushed down, twisted the locking cable to get rid of any passible film, the lead was whipped multiple times and re-inserted with no change.A new test cable was opened and the issue was resolved.This was information confirmed with the physician.
|
| |
|
Search Alerts/Recalls
|
|
|