TELEFLEX MEDICAL WECKVISTA ACCESS BALLOON PORT 12MMX70MM; LAPAROSCOPE, GENERAL & PLASTIC Back to Search Results
Catalog Number 412944
Medical Device Problem Codes Break (1069); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the balloon got burst during a urological surgery.Therefore, the device was removed and replaced with a new one to complete the procedure.Nothing fell/remained in the patient, and no injury to the patient was reported".The patient status is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review could not be performed as no lot number was reported.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.Other remarks: n/a.Corrected data: n/a.
 
Additional Manufacturer Narrative
(b)(4).Additional information received on 29 november 2023 states that "from the returned sample, the first sentence of the description is revised as follows: it was report that the ring clamp got broken during a urological surgery".The reported complaint of "broken - ring clamp" was confirmed based on the sample received.The ring clamp was observed to be broken and would not lock properly.An investigation was opened by the manufacturing site to further investigate this issue.A risk evaluation, quality alert, and supplier notification were implemented by the manufacturing site on 29-may-2023.A dhr review was performed with no evidence to suggest a manufacturing related cause.The returned sample was manufactured prior to this on 10-may-2022.Teleflex will continue to monitor and trend on complaints of this nature.Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
It was reported that "the balloon got burst during a urological surgery.Therefore, the device was removed and replaced with a new one to complete the procedure.Nothing fell/remained in the patient, and no injury to the patient was reported".The patient status is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WECKVISTA ACCESS BALLOON PORT 12MMX70MM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
mariah mackinnon
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18169495
Report Number3003898360-2023-01553
Device Sequence Number7716641
Product Code GCJ
UDI-Device Identifier04026704559763
UDI-Public04026704559763
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K112456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/09/2024
Device Catalogue Number412944
Device Lot Number73E2200360
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2023
Supplement Date Received by Manufacturer12/28/2023
Initial Report FDA Received Date11/20/2023
Supplement Report FDA Received Date01/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-